BRANDGWINSTEK
MODEL

 

A Solid Foundation for Reliable, Safe and Practical Safety Compliance Testing

Features
  • 500VA AC Test Capacity
  • 240x64 Ice Blue Dot matrix LCD
  • Sweep Function for DUT characteristic analysis
  • Insulation resistance measurement up to 50GΩ(GPT-9903/9904)
  • Manual/Auto Mode
  • Function Key for quick selection
  • High intensity flash for caution & Status indication
  • Safety INTERLOCK function
  • Zero Crossing Turn-on Operation
  • Controllable Ramp-up Time
  • True RMS Current Measurement
  • High resolution: 1uA for measuring current, 2V for setting voltage
  • PWM amplifier to enhance the power efficiency and reliable testing
  • 100 memory block for test condition (step) setting. And each step can be named individually.
  • Remote terminal on the front panel for “start” and “stop” control
  • Rear panel output available
  • Interface : RS-232C, USB device, Signal I/O and GPIB (optional)

AC WITHSTANDING

Output-Voltage Range 0.100kV~ 5.000kV ac
Output-Voltage Resolution 2V/step
Output-Voltage Accuracy ± (1% of setting +5V) [no load]
Maximum Rated Load 500 VA (5kV/100mA)
Maximum Rated Current 100mA (0.5kV < V ≦ 5kV) 10mA (0.1kV ≦ V≦ 0.5kV)
Output-Voltage Waveform Sine wave
Output-Voltage Frequency 50 Hz / 60 Hz selectable
Voltage regulation ± (1% of reading +5V) [full load → no load]
Voltmeter Accuracy ± (1% of reading+ 5V)
Current Measurement Range 0.001mA~100.0mA
Current Best Resolution 0.001mA / 0.01mA / 0.1mA
AC Current Measurement Accuracy ±(1.5% of reading +30 counts) when HISET < 1.11mA ±(1.5% of reading +3 counts) when HISET ≧ 1.11mA
Window Comparator Method Yes
ARC Detect Yes
RAMP (Ramp-Up Time) 0.1~999.9s
TIMER (Test Time)* OFF, 0.5s~999.9s
Sweep Function* Yes
GND ON/OFF

DC WITHSTANDING

Output-Voltage Range 0.100kV~ 6.000kV dc
Output-Voltage Resolution 2V/step
Output-Voltage Accuracy ± (1% of setting +5V) [no load]
Maximum Rated Load 100W (5kV/20mA)
Maximum Rated Current 20mA (0.5kV < V ≦ 6kV) 2mA (0.1kV ≦ V≦ 0.5kV)
Voltage regulation ± (1% of reading +5V) [full load → no load]
Voltmeter Accuracy ± (1% of reading+ 5V)
Current Measurement Range 0.001mA~020.0mA
Current Best Resolution 0.001mA / 0.01mA / 0.1mA
DC Current Measurement Accuracy ±(1.5% of reading +30 counts) when HISET < 1.11mA ±(1.5% of reading +3 counts) when HISET ≧ 1.11mA
Window Comparator Method Yes
ARC Detect Yes
RAMP (Ramp-Up Time) 0.1~999.9s
TIMER (Test Time)* OFF, 0.5s~999.9s
Sweep Function* Yes
GND ON/OFF

INSULATION RESISITANCE

Output-Voltage 50V~1000V dc
Output-Voltage Resolution 50V/step
Output-Voltage Accuracy ± (1% of setting +5V) [no load]
Resistance Measurement Range 0.001GW~ 50.00GW
  Test voltage Measurable Range Accuracy
  50V≦V≦450V 0.001~ 0.050GW 0.051~ 2.000GW ±(5% of reading +1 count) ±(10% of reading +1 counts)
  500V≦V≦1000V 0.001~ 0.500GW 0.501~ 9.999GW 10.00~ 50.00GW ±(5% of reading +1 count) ±(10% of reading +1 count) ±(15% of reading +1 count)
Window Comparator Method Yes
Output Impedance 600kW
RAMP (Ramp-Up Time) 0.1~999.9s
TIMER (Test Time) 1s~999.9s
Sweep Function* Yes
GND OFF (fixed)

GROUND BOND

Output-Current Range 03.00A~32.00A ac
Output-Current Resolution 0.01A
Output-Current Accuracy 3A≦I≦8A : ± (1% of setting +0.2A) 8A<I≦32A : ± (1% of setting +0.05A)
Test-Voltage 6Vac max. (open circuit)
Test-Voltage Frequency 50 Hz / 60 Hz selectable
Resistance Measurement Range 10mW~650.0mW
Resistance Measurement Resolution 0.1mW
Resistance Measurement Accuracy ± (1% of reading +2mW)
Window Comparator Method Yes
TIMER (Test Time) 0.5s~999.9s
Sweep Function* Yes
GND OFF (fixed)
Test Method Four Terminal

MEMORY

Single Step Memory MANU : 100 blocks
Automatic Testing Memory AUTO : 100 blocks, menu per auto : 16

INTERFACE

Rear Output Standard
RS232 Standard
USB (Device) Standard
GPIB Option
Remote Terminal (Front) Standard
SIGNAL IO Standard

General

DISPLY 240 x 64 dot matrix LED back light LCD
POWER SOURCE AC100V/120V/220V/230V ±10%, 50/60Hz
POWER CONSUMPTION 1000VA max.
DIMENSIONS & WEIGHT 330(W) x 148(H) x 587(D) mm (GPT-9904/9903A), 330(W) x 148(H) x 482(D) mm (GPT-9902A/9901A) approx.27kg(Max)
* The “Sweep” and “TIMER OFF” can only be performed when the tester is in the special MANU mode
Please fill out the following request form. We will contact you as soon as possible.
Your name
Your Email
Your phone
Your message
Updating

DETECH SYSTEM PTE LTD

Address: 1 Yishun Industrial St. #07-23. APosh Bizhub,
Singapore 768160